Trinocular Stereo Microscope with Camera

Trinocular Stereo Microscope with Camera

Details
Model: T45T
NOTICE
1) Microscope ought to be placed in a dry and clean place. Do not expose the microscope in the sun directly. Avoid high temperature and violent vibration.
2) As microscope is a precision instrument, handle with care, avoiding impact or abrupt movement during transportation.
3) To keep the image clear, do not leave fingerprints or stains on the surfaces of the lens.
4) Never turn the left and right focusing knob in the adverse direction at the same time, otherwise the microscope will be damaged.
5) Hold the camera with one hand for fearing of falling when you take the films out of the big camera.
Category
Stereo Zoom Trinocular Microscope
Share to
Send Inquiry
Description
Technical Parameters

Before use

 

NOTICE
1) Microscope ought to be placed in a dry and clean place. Do not expose the microscope in the sun directly. Avoid high temperature and violent vibration.
2) As microscope is a precision instrument, handle with care, avoiding impact or abrupt movement during transportation.
3) To keep the image clear, do not leave fingerprints or stains on the surfaces of the lens.
4) Never turn the left and right focusing knob in the adverse direction at the same time, otherwise the microscope will be damaged.
5) Hold the camera with one hand for fearing of falling when you take the films out of the big camera.

 

MAINTENANCE
1) All lenses must be kept clean. Fine dust on the surface of the lens should be blown off with hand blower or wiped off gently with a soft lens tissue; Fingerprints or oil marked on it should be wiped off with a tissue moistened with a small amount of xylene or a 3:7 mixture of alcohol and ether.
2) Never use the organic solution to clean the other surface (especially the plastic surfaces). If necessary, please choose the neutral detergent.
3) Do not take the microscope apart for fearing that it is damaged.
4) After using, cover the microscope with the dust-cover provided and store it in a dry and clean place free from moisture to prevent rust.
5) To keep the performance of the microscope, please check it periodically. The detail can be gotten from the agent nearby.

 

Nomenclature

 

1

2

3

 

Assemblage

 

4

5

 

Operation

 

4-1 Use the glass stage
1) Press the glass stage on the sunken place then the other side of the glass stage will be lifted. (Fig.1)
 

6

 

4-2 Adjust the degree of tightness of the focusing arm.
1) If you want to adjust degree of tightness of the focusing arm, you can hold one of the focusing knobs and turn another one to attain a suitable position. The degree of tightness relies on the direction to be turned. The clockwise direction is tight, otherwise, is loose.
2) The suitable position of tightness can make the adjustment more comfortable and prevent the focusing bracket from slipping down by its weight during the observation. (Fig.2)
 

7

 

4-3 Set the specimen slide
1) Set the specimen on the center of stage plate. If necessary, clamp the slide with the clips.
2) Turn on the light.

 

4-4 Adjust the specimen slide
1) Turn the zoom control knob to the maximum magnification.
2) Turn the diopter adjusting rings to the zero.
3) Observe the specimen through the right eyepiece and make the jkj image clear by turning the focusing knob.Fig.3
4) Rotate the zoom control knob to the minimum magnification.
5) Observe the specimen through the right eyepiece and make the image clear by turning the right diopter adjusting ring② (Fig.3)
6) Redo the step(1), (3), (4)and (5) till the right adjusting ring is more precise.
7) Do the step (4) and make the image clear which is observed through the left eyepiece by turning the left diopter adjusting ring ①. (Fig.3)

 

8

 

4-5 Adjust the interpupillary distance
1) Adjust the prism housing along the direction of arrowhead of the till the observation is comfortable. Fig.4

 

9

 

4-6 Use Eyepiece shields
1) For user who does not wear glasses, hold the diopter-adjusting ring to prevent them from rotating and turn the eyepiece till the eyepiece shields fit the observer well.
2) For user who wears glasses, take the eyepiece shields off beforeobservation

 

4-7 Mount and Remove the Optional Eyepiece Micrometer
1) Turn and remove the mounting ring ② from the eyepiece. (Fig.5)
2) Clean the eyepiece micrometer ① and mount it to the mounting ring with the inscription side downward.
3) Gently twist the mounting ring with the eyepiece micrometer into the eyepiece till tightening ② securely.
4) To remove the eyepiece micrometer, take down the mounting ring ③ by twisting and take out of the micrometer, and then wrap it.

 

10

 

4-8 Install the illumination device
1) Insert the illumination device ① in the bracket with the protrudent side toward the lock-screw ② and tighten the lock-screw. (Fig.6)
2) Put the plug into the socket of the pillar stand ③.

 

11

 

4-9 Choose the optical system
1) You can alternate the binocular observation and video capture by pushing or pulling the pole. You can attain binocular observation by pushing the pole inside, or attain video capture by pulling it outside.
No matter what optical system is chosen, push or pull the pole thoroughly.

 

12

 

4-10 Mount the photo eyepiece and the PK-mount adapter
1) Put the photo eyepieces socket of the tri-ocular.
2) Connect the PK-mount adapter with the photo eyepiece, and then tighten the lock-screw. (Fig.8)

 

13

 

4-11 Adjust the CTV
1) Adjust the CTV to a suitable position by rotating C-mount.
Note: The range of the adjustment: 1~2mm in general. (Fig.9)

 

14

 

4-12 Connect the Digital head With the Monitor or TV set
1) Plug one end of the PVA cable into the socket of the digital head. (Fig.10)
2) Plug the C-VIDEO or S-VIDEO of the PVA cable into the correct socket of Monitor (TV set).
3) Connect the 12V DC power with the power socket of the PVA cable.

 

15

 

4-13 Appear the image on the Monitor or TV
1) Connect the power supply and then turn on the Monitor or TV.
2) For the monitor, the connect sign model must be chosen (C-video or S-video) and for TV, the channel must be set to the video channel.
3) Pull the pole out and adjust the focusing knob and then the image will appear on the screen clearly.

 

4-14 Connect with the computer
1) Plug one end of the PVA cable into the socket of the digital head.
2) Plug one of the C-VIDEO or S-VIDEO into the A/D board.
3) Plug the USB of the A/D board into the USB socket of the computer. (Fig.11)
4) If your computer is mounted capture card, you can connect the C-VIDEO or S-VIDEO with the computer directly.
5) Connect the 12V DC power with the power socket of the PVA cable.

 

16

 

4-15 Appear the image on the computer
1) Turn on the power supply and let the computer work.
2) Install the software and the driver of the A/D board. (If they have been installed, this step can be omitted.
3) Double click the icon of the software, and then the video window will appear. You can set the size of the window according to your linking
4) Draw out the pole and adjust the focusing knob, and then the image will appear on the computer screen clearly.
5) If no image or the image without color, it may be because the model of the input signal does not match the output signal of CCD or the model of C-VIDEO/S-VIDEO is not correct. The detail of operation refers to《Software operation manual》.

 

4-16 Appear the image on the computer and the Monitor synchronously
1) Do step 4-12 and step 4-14 to connect the computer and the Monitor.
2) Operate step 4-13 and step 4-15, we can make the image appear on the computer and Monitor at the same time.

 

4-17 Adjust the image
1) Put the base, stand and digital head correctly, then fix the lock-screw tightly.
2) Put the object on the base stage.
3) Observe the object through the eyepiece and adjust the focusing knob to make the image of the object clearly.
4) Move the digital head or the object gently to adjust the image agreeing with observer.

 

4-18 Brief instruction for the software
1) The program design of the software is up to date, and the Chinese/English interface can berth powerful delineation bar which be used much conveniently and rapidly. You can finish most of analyze work only to click the mouse.
2) Can afford many powerful area choosing tools which can analyse any area your linking at will, such as adjusting hue and image, dealing with mathematical morphology, image matching, texture analyse, character identify and so on.
3) Geometry character measuring function, automatically analyzing function such as slightness body, grain body, line body and so on. The outcome can be kept in data and can be made into chart and so forth.

 

4-19 Use the white balance
1) The CCD has auto white balance when the white balance switch is on "ON".
2) Please put the switch on "ON" in general. Let the switch be "OFF" only in special, for example, observing the red cell, otherwise the color of red cell will be adjusted into white.
3) If you want to observe another single color, please let the switch be"ON"again when you finish the observation, and put the switch on "OFF" again after auto balance, or the color of the image will be distortion. (Fig.12)

 

17

 

4-20 Adjust the brightness of the bottom light
1) Turn the adjustable light knob① according to the sign marked on the base, along the clockwise the brightness will be added, otherwise it will be weakened. (Fig.13)

 

18

 

4-21 Replace the lamps
1) Press the stage on the sunken place then the other side will be lifted. (Fig.14)

 

19

2) Take the lamp out of the jack.
3) Put a new lamp into the jack thoroughly.
4) Recover the stage plate. (Fig.15)
Note: ① Before replacing the lamps, turn off the power first. ② Avoid violence while the lamp is plugged into the jack.

 

20

 

4-22 Replace the fuse
1) Screw the fuse tube out with a screwdriver and then pull the fuse out of the tube①.
2) Renew the fuse and mount it in an adverse way. (Fig.16)

 

21

 

Configuration

 

5-1 SZM series configuration

Configuration

Model

 

Parts

Specification

SZM45B1

SZM45B2

SZM45B3

SZM45T1

SZM45T2

SZM45T3

SZM45V2

SZM45P2

Eyepieces

SZMEWh10X20

O

O

O

O

O

O

O

O

SZMEWh15X15

               

SZMEHWh20X10

               

Binocular

SZM7045

O

O

O

         

Tri-ocular

SZM7045TR

     

O

O

O

   

Digital head

SZM7045V

           

O

O

A/D board

W-A/D

             

O

PAV cable

W-RVB-PAV

           

O

O

Power adaptor

W-BW230-12-4.8

           

O

O

Software

Pholib3.0

             

O

Auxiliary objective

SZMAO0.5/165mm

               

SZMAO1.5/45mm

               

SZMAO2/30mm

               

Focusing arm

SZMA1

O

O

O

O

O

O

O

O

Stand

SZMST1

O

   

O

       

SZMST2

 

O

   

O

 

O

O

SZMST3

   

O

   

O

   

SZSTL1

               

SZSTL2

               

Transformer

SZT1

               

Epi-illuminator

SZML1

 

O

O

 

O

O

O

O

Hold for illuminator

SZFH1

               

SZPD1

               

Photo device

SZMPH

               

TV adapter

SZMCTV

               

Gem clamp

S/ST-GC

               

Dark field condenser Light

ST-30-2L-HJ-01

               

Ring fluorescence

SZRL

O

   

O

       

Box

Inside foam

Outside carton

O

O

O

O

O

O

O

O

Note: The items marked"O"included and others for option

 

22

 

Technical parameter

 

6-1 SZM7045/SZM7045TR

Eyepiece

Standard configuration

Auxiliary objectives

0.5X

1.5X

2X

Working distance100mm

Working distance165mm

Working distance 45mm

Working distance 30mm

Magnification

Field of view

Magnification

Field of view

Magnification

Field of view

Magnification

Field of view

10X/20

7X

28.6

3.5X

57.1

10.5X

19

14X

14.3

45X

4.4

22.5X

8.9

67.5X

3

90X

2.2

15X/15

10.5X

21.4

5.25X

42.8

15.75X

14.3

21X

10.7

67.5X

3.3

33.75X

6.7

101.25X

2.2

135X

1.7

20X/10

14X

14.3

7X

28.6

21X

9.5

28X

7.1

90X

2.2

45X

4.4

135X

1.5

180X

1.1

 

6-2 The base electronic specification of SZM series

Model Parts

SZMST1

SZMST2

SZMST3

Power supply

No

220V-50Hz,110V-50/60Hz

220V-50Hz,110V-50/60Hz

Transformer

No

Input: 220/110VAC

Output: 12V DC/45W

Input: 220/110VAC

Output: 12V DC/45W

Illuminator

Top light

No

12V/15W halogen lamp

12V/15W halogen lamp

Bottom light

12V/15W halogen lamp

220/110V,7W fluorescent lamp

 

6-3 CCD Specification

CCD Model Specification

NTSC

PAL

Pick-up Device

1/3 ″

1/3 ″

Vertical Resolution

450 TV line

450 TV line

Number of Pixels

(H)768*(V)494

(H)712*(V)582

Scanning System

525 lines, 60Field/Second

625lines,50 Field/Second

White Balance

Can be switched between auto white balance and hand white balance

Back light compensation

Auto

Auto

Signal/Noise ratio

More than 46db

More than 46db

Gamma Characteristic

0.45

0.45

Minimum illumination

3 Lux

3 Lux

Input voltage

12V DC(9V-14V)

12V DC(9V-14V)

Power consumption

1.85 or less

1.85W or less

 

★ Working distance is fixed regardless of the magnification factor.
★ Total mag.= Zoom mag. X Eyepiece mag. X Auxiliary objective mag.
Diameter of field of view(mm)=Field number of eyepiece/ Zoom mag.X Auxiliary objective mag.
★ Photo adaptor mag.= Zoom mag.( ×Auxiliary objective mag.)×Eyepiece mag.
★ TV adaptor mag.=Zoom mag.(Xauxiliary objective mag.)X C-mount TV adaptor middle
★ Field of video view is 83%
★ Total video magnification range is 18∽117

 

Troubleshooting

 

The performance of the microscope can't be made fully because of unfamiliar using, this table will give some advices.
7-1 General troubleshooting

Trouble

Cause

Remedy

Double images

Interpupillary distance is not correct

Readjust it

Diopter adjustment is not correct

Readjust it

Magnification of each eyepiece is not the same size

Mount the same size eyepiece

Dirt appears in the field of view

Dirt on the specimen

Clean the specimen

Dirt on the surfaces of eyepiece

Clean the surface

Image is not clear

Dirt on the surfaces of the objectives

Clean the objectives

Image is not clear while the focus changing

Diopter adjustment is not correct

Readjust the diopter

Focus is not correct

Readjust the focus

The focusing knob is not smooth

The focusing knob is too tight

Loosen it to a suitable position

The image is obscure because of the head slipping down by itself during observation

The focusing knob is too loose

Tighten it to a suitable position

Incision image appears in the field of view or of the video view

The pole is not in correct position

Pull or push it to the correct position

The image on the monitor is not clear when the focusing knob is turned.

The focus of video is not correct

Readjust the focus of video to a correct position

Eyes fell tired easily

Diopter adjustment is not correct

Adjust the diopter

Brightness of light is not correct

Adjust the brightness

Bulb does not work when the switch is on

No power supply

Check the connection with the power supply

The bulb was not inserted correctly

Insert it correctly

Bulb is wrong

Replace with a new one

Bulb is burned out suddenly

Use the wrong bulb

Replace with a correct one

The voltage is too high

Control the voltage

Eg: use voltage regulator

Brightness is not enough

Use a wrong bulb

Replace with a correct one

The voltage is too low

Increase the input voltage

The bulb flickers or the brightness is unstable

The bulb will burn out soon

Replace with a new one

The bulb was not inserted correctly

Insert it correctly

 

7-2 Video troubleshooting

Trouble

Cause

Remedy

Incision image appears in the video view

The pole is not in correct position

Draw it to the correct position

Dirt appears in the video view

Dirt on the specimen

Clean the specimen

Dirt on the surface of objective

Clean the surface

Image is not clear while the focus changing

The image is not clear in the high magnification

Readjust the high magnification

No image on the TV screen

The draw pole is not in correct position

Draw it to the correct position

Objective cover is not open

Open it

TV is not on Video channel

Choose the correct one

No image on the Monitor

Connection is not correct

Reconnect the circuit

Objective cover is not open

Open it

The input signal does not accord with the signal be chosen on the Monitor

Choose the correct signal model

The software run slowly or the window of the view does not come out

12V DC power does not be connect

Connect the 12V DC power

No input signal of A/D board

Reconnect the C-Video or S-Video signal

The input signal does not accord with the signal which is chosen in the driver of the A/D board

Choose the correct signal model which match the input signal

The image is not correct on the view window

The CCD model chosen in the driver of the A/D board does not accord with the real CCD

Choose the correct CCD model

 

 

Hot Tags: trinocular stereo microscope with camera, China trinocular stereo microscope with camera manufacturers, Zoom Stereo Microscope, 3 5 X 90x Trinocular Stereo Microscope, Stereo Zoom Microscope with Digital Camera, stereo zoom trinocular microscope, Trinocular Stereo Microscope 7 45X with Camera, Trinocular Stereo Microscope with Camera